Medical Nebulizer Inhaler Children Adult Household Medical Air Compressed Nebulizer
Usage
1.For type,dose,and regime of medication follow the instructions of your doctor or respiratory therapist.
2.Clean and disinfect the nebulizer kit,mouthpiece and optional nosepiece or masks before using them for the first time after purchase,if the device has not been used fora long period of time.
3.If the device is used by more than one person,e sure to clean and disinfect parts before they are used by each person.
4.Always dispose of any remaining medication after use,and use fresh medication each time.
5.Do not use only water in the nebulizer,for nebulizing purposes.
6.Do not use the device where it may be exposed to flammable gas.
7.Do not block the ventilation slots.Never place the device where the ventilation slots may be obstructed during operation.Do not cover the compressor with a blanket or towele tc.,duringuse.
Operation Environment
|
Temperature: 5ºC~40ºC.
|
||||||
Relative humidity: ≤80%.
|
|||||||
Atmosphere pressure: 80.0kPa~106.0kPa
|
|||||||
Spray rate
|
The spray rate should be ≥0.2mL/min.
|
||||||
Air flow volume
|
Compressed air flow of compression pump:≥8L/min.
|
||||||
Pressure range
|
1. Under normal working conditions, the pressure range of device is 60kPa~130kPa. 2. When the pipeline blockage occurs etc. in the device, the pressure produced by the device will not break in the range of 130kPa~ 400kPa. |
||||||
Residual liquid volume
|
≤0.7mL
|
||||||
Machine noise
|
Noise of machine in normal operation: ≤ 65dB (frequency-weighting characteristic A).
|
||||||
Continuous working hours
|
When the device works continuously for more than 4 hours under normal temperature and AC power supply, it should be able to work normally.
|
||||||
Enclosure leakage current
|
Normal condition≤0.1mA; Single fault condition≤0.5mA.
|
||||||
Patient leakage current
|
Nomal condition≤0.1mA; Single fault condition≤0.5mA.
|
||||||
Droplet particle size
|
1um-5um
|