Customization: | Available |
---|---|
Type: | Vet Anesthesia Machine |
Application: | Orthopedic |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Physical specifications
|
|
Screen:
|
5.4'' Display Screen
|
Suitable
|
Adult & Child
|
Mode:
|
pneumatically powered and electrically controlled system
|
Working Mode:
|
Closed; Semi-Closed; Semi-Open
|
Circuit
|
Breathing circuit integrated standards
|
Tube:
|
4Tubes Flowmeters: O2:0.1~10L/Min, N2O:0.1~10L/Min
|
Trolley:
|
Fitted with 4 nos anti-static rubber castors; two of which are lockable for braking and easy maneuverability with foot operated
brake provisions |
Drawer unit
|
Two drawer totally extractable
|
Technical Specifications:
|
|
Gas requirement:
|
Medical oxygen and nitrous oxide with a pressure ranging from O2: 0.32~0.6MPa ; NO2: 0.32 MPa to 0.6 MPa.
|
safety valve
|
<12.5 kPa
|
Respiratory Rate
|
1~99bpm
|
Flow meter
|
O2:0.1~10L/Min, N2O:0.1~10L/Min
|
oxygen concentration in the mixed gas N2O/O2
|
> 21%
|
Oxygen Flush:
|
25~75 L/min
|
modes of ventilation
|
A/C, IPPV, SIPPV, IMV, SIMV, MANUAL
|
I/E Ratio:
|
4:1 ~ 1:10, Has the inverse ratio ventilation
|
Tidal Volume
|
0~1500ML
|
Inspiratory trigger pressure
|
-0.1kPa~1.0kPa
|
Inspiratory Plateau:
|
0~1s
|
O2 Concentration:
|
21%~100%
|
Maximum safety pressure:
|
≤ 12.5 kPa
|
Pressure limit range:
|
0 ~ 6.0 kPa
|
Airway pressure alarm: Audible and visual and with yellow and red color indicating
|
Lower: 0.2kPa~5.0kPa; Upper: 0.3~6.0 kPa
|
±0.2 kPa
|
|
Tidal volume alarm: Audible and visual and with yellow and red color indicating
|
upper alarm: 50 to 2000ml, lower alarm: 0~1800ml
|
Power Supply Alarm
|
Ac/dc power supply are after failing to send out alarm immediately
Alarm Time: keep >120s |
The airway pressure continues to be higher than 15 hPa ±1 hPa for 15s±1s, then the machine will raise an audible alarm, the
pressure will be displayed in red and the continuous high pressure red alarm wording is displayed on the screen of the anesthetic respirator. |
|
Operating conditions
|
|
Ambient temperature:
|
10 ~ 40oC
|
Relative humidity:
|
no higher than 80 %
|
Atmospheric pressure:
|
860 hPa ~ 1060 hPa
|
Power requirement:
|
100-120 Vac, 50/60 Hz;
|
Air source requirement:
|
Medical oxygenwith a rated pressure ranging from 0.3 to 0.5MPa.
|
Attention: the AC power supply used for the anesthesia machine must be well grounded.
|
|
Attention: the anesthesia machine used must be equipped with a carbon dioxide monitor complying with ISO 9918:1993, an oxygen
monitor complying with ISO 7767:1997 and an expiratory gas volume monitor complying with 51.101.4.2 of Medical Electrical Equipment Part II: Special Requirements for Safety and Basic Performance of Anesthesia System. |
|
Storage
|
|
Ambient temperature:
|
-15oC ~ +50oC
|
Relative humidity:
|
no higher than 95 %
|
Atmospheric pressure:
|
86 kPa ~ 106 kPa.
|
|
It should be stored in a room without corrosive gas and well ventilated
|
Package
|
|
packaging box
|
comply with the requirement of GB/T 15464
|
|
Between the packaging box and product, soft material with appropriate thickness provided to prevent loosening and mutual friction
during transportation |
|
Moisture protection and rain protection to ensure that product is protected from natural damage.
|
Safty & Alarm
|
|
Oxygen Alarm
|
It alarms when the oxygen supply from pipe or cylinders lower than 0.2MPa
|
Ventilation Volume Alarm
|
Lower: 0~12L/Min; Up: 18L/ Min
|
Power Alarm
|
It Alrmas while AC and DC supply failure ; Keep alraming time: >120s
|
Air Tract Pressure Alarm
|
Lower: 0.2kPa ~5.0 kPa; Up: 0.3kPa ~ 6.0kPa
|