Customization: | Available |
---|---|
Classification: | Imaging Diagnostic Equipment |
Type: | X Ray Equipment |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Our MCI0500 3d digital x ray mammography machine is the digital mammography specialists, it is used to the digital diagnostic mammogram, MeCan Medical is the best superior digital x ray mammography manufacturer near me.
1.Pitch function
2.Point-Pressure Magnification Photography
3.Tungsten target tube reduces 30%~60% radiation exposure.
4.AI assisted mammography diagnosis system
5.Amorphous selenium detector: the groups of calcification points showing increases from 3 to 6.
6. With wheelchair mode
7. Tube Type: Molybdenum
Digital mammography specification
No. | Parameters and performance | |
1 | Basic function | |
1.1 | Application: it is specially used for human breast tissue photography to obtain tissue images for breast disease diagnosis and breast cancer screening; | |
2 | Rack system | |
2.1 | Frame structure: vertical C-arm type, | |
2.2 | C-arm rotation range + 195 ° to - 155 ° | |
2.3 | Vertical movement range of C-arm: 630mm-1440mm | |
2.4 | The C-arm can be locked at any position in the whole stroke | |
2.5 | The rack display screen can display the current position and position information to be photographed, and has the function of one key in place, which can switch the preset position with one key |
|
2.6 | Distance between focus and image receiving surface (FID): 660mm | |
3 | High voltage generator assembly (imported) | |
3.1 | Maximum output power: 5KW | |
3.2 | Output voltage range: 20-40kv | |
3.3 | Output current range: 10ma-160ma | |
3.4 | Loading time range: 0.005s-10s | |
3.5 | Current time product regulation range: 0.5-560mas | |
3.6 | With exposure control system, support one key exposure or hand brake exposure. | |
4 | X-ray tube assembly (imported) | |
4.1 | Double speed rotating anode | |
4.2 | Nominal tube voltage: 40kV | |
4.3 | Anode heat capacity: 300KHU | |
4.4 | Anode rotating speed: 10000 rpm | |
4.5 | Anode material: Molybdenum | |
4.6 | Nominal focus size: small focus 0.1mm, large focus 0.3mm | |
5 | Flat panel detector Component | |
5.1 | Imaging medium: amorphous silicon | |
5.2 | Imaging size: 240mm × 300mm | |
5.3 | Conventional spatial resolution: 7.0lp/mm | |
5.4 | Pixel matrix: 4096*3072 |