Customization: | Available |
---|---|
Certification: | CE, ISO13485 |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Channel
|
Single
|
Stackable
|
YES
|
Syringe Size
|
5,10,20,30,50/60ml
|
Infusion Modes
|
Rate Mode, Rate- Time, Rate-VTBI ,Time-VTBI, Body Weight
|
Safety
|
Double CPU to make sure higher standard infusion
|
Drug Library
|
20 drug list with drug code display
|
Accuracy
|
±2%
|
Time Preset
|
00:01-99: 59 (Hour: Min)
|
Volume Range
|
0-9999.9ml
|
Flow Rate Step by
|
0.1 m/h when Rate<100ml/h, 1ml/h when Rate >=100ml/h
|
Flow Rate Range
|
5ml Syringe 0.1m/h-150ml/h
|
10ml Syringe 0.1m/h-300ml/h
|
|
30ml Syringe 0.1m1/h-900mlh
|
|
50/60ml Syringe 0.1m/h-1200m/h
|
|
Purge/Bolus
Rate/Top Flow Rate
|
5ml Syringe 150m/h ; 10ml Syinge 300ml/h; 20ml Syringe 600m/h;
30ml Syringe 900mlh; 50/60ml Syringe 1200ml/h
|
Audible & Visible Alarms
|
Auto Sef-Test, Syringe Dislocation, Occlusion, Near End,Syringe Empty, VTBI Completion, Low Battery, Battery Exausted, Motor Malfunction, Incorrect Syringe Specification, Circuit Malfunction,
Master CPU Malfunction, Monitoring CPU Malfunction, Parameters Over Limitation, A.C Disconnection, A.C connection
|
KVO
|
0.1-5.0ml/h Adjustable
|
Occlusion Pressure
|
High:40 KPa+20KPa
Middle:60 KPa+20KPa
Low:100KPa+20KPa
|
Max Infusion Pressure
|
120KPa
|
Battery
|
>=8 hours
|
Power Consumption
|
30VA
|
Power Supply
|
AC: 100-240V, 50Hz/60Hz
|
Battery
|
Lithium bttery.11.1/2000mAh
|
Classification
|
Class I,type CF,IPX4
|
Dimension
|
26x21.5x11cm
|
Weight
|
2 kg
|
Optional Function
|
Ambulance DC: 12V
|