• Dual Beam Infrared Spectrophotometer
  • Dual Beam Infrared Spectrophotometer

Dual Beam Infrared Spectrophotometer

Application: Industrial
Certification: ISO9001, CE
Customized: Non-Customized
Product Name: Dual Beam Infrared Spectrophotometer
Transmittance (%): 0-100.0%
Absorbency: 0-3ABS
Customization:
Diamond Member Since 2015

Suppliers with verified business licenses

Trading Company

Basic Info.

Model NO.
MCL-SP06
Wavenumber Range
4000-400 Cm-1
Transport Package
Standard Export Package
Trademark
MeCan
Origin
China (Mainland)
Production Capacity
1000

Product Description

Dual Beam Infrared Spectrophotometer

MCL-SP06

What is feature of our Infrared Spectrophotometer?

Description
Infrared spectroscopy is one of the most powerful techniques for identifying organic substances in organic and analytical chemistry. Infrared analysis can be both qualitative and quantitative. MCL-SP06 is an important tool in analytical laboratories.

MCL-SP06 Dual-Beam Infrared Spectrophotometer can be used to record the IR absorption and reflectance spectra of substances in the spectral range of 4000 ~ 400 cm-1. It is a powerful tool to analyze sample structures in the fields such as petroleum, chemical engineering, pharmacy, public health and environmental protection.

The application software provides a user-friendly interface for the control of the spectrophotometer, data acquisition, and spectral analysis with functions listed below:

-Spectral background baseline memory

-Spectral background baseline correction

-Spectral data smoothing operation

-Spectral baseline slope correction

-Spectral data differential operation

-Spectral data arithmetic operation

-Spectral data accumulating operation

-Conversion of %T and Abs

-Spectrum file management

-Spectral peak search

-Spectrum scale extension

-Spectral absorption expansion

What is specification of our Infrared Spectrophotometer?

Wavenumber Range

4000-400 cm-1

Transmittance (%)

0-100.0%

Absorbency

0-3Abs

Wave-number Accuracy

≤±4 (4000-2000cm-1)

≤±2 (2000-500 )

WN Repeatability

≤2 (4000-2000cm-1)

≤1 (2000-450cm-1)

Transmittance Accuracy

≤±0.5%(noise level not included)

Transmittance Repeatability

≤0.5%(1000-930cm-1 )

Line Flatness and Straightness

≤4%

Resolution Capability

Polystyrene has six absorption peaks around 3000 ,with height of 1% at least; Ammonia gas's resolution is 2.5 around 1000 , with height of 1% at least.

Stray Lights

≤1%(4000-650cm-1 )≤2%(650-400cm-1 )