Customization: | Available |
---|---|
Weight: | 2.48 Kg |
Screen: | 4.3 Inch LCD Screen |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
a) Infusion Rate range(mL/h) | 1-1200; 1mL/h~99.9mL/h range adjustable in 0.1mL/h steps or 1mL/h steps; 100.0mL/h~1200mL/h adjustable in 1mL/h steps only; |
b) Infusion precision | ±5% (correctly calibrated and operated in a stable environment according to the requirement of 10) |
c) Pre-set of infusion volume(mL) | 0~9999, 1mL steps |
d) Infusion volume error | ±5%; (correctly calibrated and operated in a stable environment according to the requirement of 10) |
e) Bolus rate(mL/h)and error | 200~1200 continuously adjustable;±20%; |
f) Occlusion alarm level(pressure) | 50kPa±25kPa |
g) Max. pressure from the infusion set | More than 120kPa |
h)Triggering time of occlusion alarm and bolus dose | Minimum flow rate:When the occlusion alarm pressure is 50kPa±25kPa, Less than 13 minutes. Intermediate flow rate: When the occlusion alarm pressure is 50kPa±25kPa, Less than 14 seconds. Occlusion dose does not beyond 0.3mL (use Jerry infusion set, chokepoint is made at the tail end of the infusion set.) |
i) Keep vein open(KVO)rate | KVO rate of 3mL/h with flow rate higher than 10mL/h,1mL/h with flow rate lower or equals to 10mL/h and higher than 1 mL/h,0.1 mL/h with flow rate lower than 1 mL/h, accuracy error+ 20% |
j)Recoverable alarm | 1 min 50 s ~ 2 min |
k) Pause overtime alarm time | 1 min 50 s ~ 2 min |
l)Classification | Type CF, Class II continuous volumetric infusion pump with internal battery; IPX4 (sealed drip-proof device) |
m) environmental temperature for rechargeable battery | environmental temperature: -20ºC~+60ºC |
n) service life | 5years |