The MCI0199 system is a medical diagnostic magnetic resonance system. The main magnet of the system adopts the "C" type, single pole structure. The structure has large open degree (the opening angle is larger than 270 degree), which provides the patient much comfort and acceptance. The main magnet has the advantage of compact structure, small size (1.9 m×1.3 m×1.8m) and light weight (16T) with the static magnetic field strength up to 0.4 Tesla.
System Siting Requirements
Power Voltage Regulator will be supplied for the system by WDM.
RF shielding solutions and site planning support is provided through WDM.
Magnet room requires 24hr isolate Air Conditioning to stabilize the room temperature at 23 centi-degree with deviation of ±1 centi-degree per hour. Humidity: 40 - 60%.
Electronics room requires air-conditioning, room temperature: 15 - 30 centidegree with deviation less than 3 centidegree per hour. Humidity: 40 ~ 80%.
Air Conditioning Machines are to be provided by customer, and WDM is available to supply Air Conditioning Machines upon customers request at additional cost to the customer. Specifications of Air Conditioning Machines depend on size of magnet room. Normally it is suggested for the magnet room to have refrigerating capacity no less than 3KW and able to work 24 hrs x 7 days